
1100 LC Modules Control Manual - 59 - 03-914947-30:Rev.6
Pressure Alarms Parameter Group
Enter the Minimum and Maximum pressure limits of the pump.
Outside these limits, the pump is stopped.
Flow Alarms Parameter Group
Enter the Maximum Flow Ramp of the pump. Outside this limit,
the pump is stopped.
Control Control
Primary Channel Parameter Group
Select in the scrolling list the channel where the following actions
will be done. The available choices are: Automatic, Channel A,
Channel B, Channel C, Channel D.
Comentarios a estos manuales