1100 LC Modules Control Manual - 174 - 03-914947-30:Rev.6
For analyzing the test results, refer to the detector operating
manual.
Wavelength Calibration Test
Perform the Wavelength Calibration test to check wavelength
calibration.
During the test, the alpha (656.1 nm) and beta (486 nm)
emission lines deviation of the deuterium lamp are evaluated.
The test should be done with water in the flow cell.
The 1 nm slit is moved into the light path and the gain is set to
zero. The largest emission lines deviation is displayed.
At stage 6, choose to use this result to recalibrate the detector by
clicking Yes. If the deviation is small, click on No to ignore the
results.
At the end of the re-calibration, the largest emission lines
deviation is evaluated again and displayed on screen and also in
the maintenance audit trail:
For analyzing the test results, refer to the detector operating
manual.
D/A Converter Test
Perform the D/A Converter test to check correct operation of the
digital-analog-converter.