138 1100 Series Capillary LC System Manual
6Options
Installing the Extended Flow Range Kit
1 Using the 4 mm / 1/4 - 5/16 inch open wrenches disconnect the capillaries
8, 9, 10, 11 and 13. Refer to Figure 39 on page 137 to identify their location.
2 Remove the 20 µl flow sensor by unscrewing the 2 holding screws with the
2.5 mm hex key.
3 Install the 100 ul flow sensor and screw the 2 holding screws with the
2.5 mm hex key.
4 Using the 4 mm / 1/4 - 5/16 inch open wrenches connect the capillaries 8, 9,
10, 11 and 13 (refer to Figure 39 on page 137) to identify their location
.
.
Frequency When the flow rate is higher than 20 µl/min
Tools re quired 4 mm open wrench (8710-1534)
Torque adapter G1315-45003
1/4 - 5/16 inch open wrench (8710-0510)
2.5 mm hex key (8710-2412)
Parts required Extended flow range kit (G1376-69707)
If the pressure drop in the system is not to high you can leave the capillary G1375-68703
between the column and the detector (item 8). Otherwise change it as recommended with
the capillary G1375-87308.
To successfully install the capillaries and avoid the risk of a leak go to Chapter 3,
“Capillaries and Fittings.