44 1100 Series Capillary LC System Manual
2 Optimizing Performance
2 When the Fast System Flush time has elapsed, the micro well-plate sampler
needle is returned to the needle-seat. The pump returns to the normal
operating mode, reconditioning the column at the flow and initial
composition defined in the current method. The column is reconditioned
for the Column Reconditioning time defined in the user interface.
If multiple injections are in progress, the next injection will begin when Fast
Composition Change/Reconditioning is completed.