1100 Series Capillary LC System Manual 23
Installing your Capillary LC System 1
Manually Priming the Solvent Channels.
1 The degasser accessory kit contains a 20ml plastic syringe and a solvent
tube adapter for this syringe. Push the adapter onto the syringe.
2 Pour the intended analytical solvents into the solvent bottles, and install
the bottles on the desired solvent channels. Install Isopropanol on channels
which will not be used right away.
3 Put a paper towel over the leak sensor in the pump leak tray.
4 Disconnect the channel A solvent tube from the A1 port of the pump solvent
selection valve.
5 Connect the end of the solvent tube to the syringe adapter. Slowly draw a
syringe volume (20ml) from the solvent tube.
6 Disconnect the solvent tube from the syringe adapter, and reconnect the
tube to the A1 port of the solvent selection valve. Eject the syringe contents
into an appropriate waste container.
7 Repeat steps 4 to 6 for the three remaining solvent channels.
8 When all 4 solvent channels are manually primed, remove the paper towel
from the pump leak tray. Make sure that the pump leak sensor is dry before
turning on the pump.
NOTE
This procedure should be done before the modules are turned on.
WARNING
Liquid may drip from the disconnected solvent tube. Make sure to follow appropriate
safety precautions.