1100 Series Capillary LC System Manual 147
Options 6
Removing the Micro Column Switching Valve
When required If valve failed or bottom foam part has to be removed for other replacements
Tools re quired Screwdriver Pozidriv 1 PT3
Wrench 5.5 mm for capillary connections
Preparations for this procedure:
Turn off the column compartment.
Disconnect the power cable.
Disconnect capillaries.
Remove column compartment from stack and place it
on the working bench.
Remove the front cover, top cover and top foam
section.
1 Disconnect the grounding connection of the valve
at the Z-panel.
2 Unscrew the Z-panel. 3 Press against the rear of the Z-panel to release the
metal plate from the guide and pull it carefully
upwards.