Agilent Technologies 1100 Series Manual de instrucciones Pagina 49

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 196
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 48
1100 Series Capillary LC System Manual 39
Optimizing Performance 2
Fast Gradients range (800-1300 µl/min)
In this range, the pump gradient delay time is as short as possible. This range
is specifically recommended for fast-gradient analyses (<3 min.). Solvent
consumption is highest in this range.
Table 5 gives approximate primary flow values (in µl/min) as a function of
selected primary flow range vs. system pressure:
Actual primary flow values may vary from system to system. In any case the
standard configuration is changed, the primary flow could be higher
compared to the values in above table.
Tabl e 5 Primary flow overview for standard pump configuration
0bar
System
pressure
100 bar
System
pressure
200 bar
System
pressure
300 bar
System
pressure
400 bar
System
pressure
Low consumption
range
200 225 250 275 300
Default range 500 570 640 710 780
Fast gradient range 800 995 1190 385 1580
Vista de pagina 48
1 2 ... 44 45 46 47 48 49 50 51 52 53 54 ... 195 196

Comentarios a estos manuales

Sin comentarios