1100 Series Capillary LC System Manual 57
Basic System Troubleshooting 4
System Pressure Abnormally High
Symptoms:
The current system pressure is significantly above the typical system pressure
produced by this method with this column.
System Pressure Abnormally High: Possible causes and actions
Possible Causes Suggested Actions Notes
The analytical column has
become plugged
Replace the column. Otherwise, backflush the
column or replace the column inlet frit.
The filter in front (upstream)
of the EMPV has become
plugged
Perform a pump purge at 1000 µl/min using
pure water. During the purge, check system
pressure. If pressure is >10 bar, replace the
EMPV filter.
If a new filter does not reduce the
pressure, replace the mixer.
A component in the
micro-sampler has become
plugged. This could be the
sample loop, needle, needle
seat assembly or injection
valve ports.
Using the sampler maintenance positions,
switch the sampler injection valve from
mainpass to bypass. If pressure is significantly
reduced:
a Backflush or change the needle seat assembly.
b Replace the needle.
c Backflush or replace the sample loop capillary.
d Replace the injection valve rotor seal.
e Clean the stator head with acetone, and make
sure the stator head ports are free of particles.
For severe capillary tube plugs,
acetone is a good backflushing
solvent.
A pre-sampler, or
post-sampler capillary in the
system has become plugged,
broken, crushed by a module
cover or overtightened.
Refer to the system flow diagram. One at a
time, disconnect the capillaries in the following
order. When the defective capillary is found, it
may be backflushed with acetone, or replaced.
a EMPV-to-flow sensor capillary
b flow sensor-to-sampler injection valve capillary
c sampler injection valve-to-column inlet
capillary
d flow cell assembly (includes inlet and outlet
capillaries)