166 1100 Series Capillary LC System Manual
7 Performance Specifications
Performance Specifications Agilent 1100 Series Capillary Pump
Tabl e 86 Performance Specification Agilent 1100 Series Capillary LC System
Type Specification
System delay volume Typically 5 µl from EFC to column head, for flow rates up to 20 µl/min
(default setup).
Typically 14 µl from EFC to column head, for flow rates up to 100 µl/min
(default setup).
Tabl e 87 Performance Specification Agilent 1100 Series Capillary Pump
Type Specification
Hydraulic system Two dual piston in series, with proprietary servo-controlled variable
stroke drive, floating piston, active inlet valve, solvent selection valve
and electronic flow control for flow rates up to 100 µl/min
Settable column flow
range
0.01 – 20 µl/min
0.01 – 100 µl/min (with the extended flow range kit)
0.001 – 2.5 µl/min (with the electronic flow control bypassed)
Recommended column
flow range
1–20µl/min
10 – 100 µl/min (with extended flow range kit)
0.1 – 2.5 ml/min (with the electronic flow sensor bypassed)
Column flow precision < 0.7 % RSD or 0.03 % SD (typically 0.4 % RSD or 0.02 % SD), at
10 µl/min and 50 µl/min column flow (based on RT, default setting)
Optimum composition
range
1 to 99% or 5 µl/min per channel (primary flow), whatever is greater
Composition precision < 0.2 % SD, at 10 µl/min (20 µl flow sensor), 50 µl/min (100 µl flow
sensor) and 1 ml/min (normal mode) default setting