1100 Series Capillary LC System Manual 49
Capillaries and Fittings 3
SST: Stainless steel
* * PFS: Peek coated fused silica
Descriptions
to Table 1 0
1 The capillary G1375-87400 connects the mixer and the manual purge valve.
2 The capillary G1375-87318 connects the manual purge valve and the injection valve (port1).
3 The capillary G1375-87312 connects the injection valve (port 6) and the heat exchanger (IN).
4 The capillary G1375-87306 connects between the injection valve (port 6) and the heat exchanger
(IN) if the thermostat G1330A/B is installed.
5 The capillary G1316-87300 connects between the heat exchanger (OUT) and the column.
Tabl e 10 Specific capillaries for use with a flow higher than 200 µl/min
Item
descriptions see
list below)
Fitting
type
*
* see Ta b l e 1 4 on page 52
Diameter
(µm)
Length
(mm)
Material Volume
(µl)
Pressure
drop (Bar)
Part number
see description 1
below
A/A 170 280 SST
6.4 2 G1375-87400
see description 2 B/C 125 550 PFS** 6.8 15 G1375-87318
13 B/C 100 200 PFS 1.6 13 G1375-87312
14 (micro ALS) B/B 250 1800 SST 88 3 G1329-87302
14 (micro WPS) B/B 250 1800 SST 88 3 G1377-87300
see description 3 B/C 100 200 PFS 1.6 13 G1375-87312
see description 4 C/B 100 550 PFS 4.4 37 G1375-87306
see description 5 A/A 170 70 SST 1.6 <1 G1316-87300
11 A/A 170 380 SST 8.6 3 G1315-87311
NOTE
The pressure drops in Ta b l e 1 0 are calculated for water (viscosity 1) and for a flow rate of
1000 µl/min.