1100 Series Capillary LC System Manual 35
Optimizing Performance 2
Cleaning the Solvent Filters
Remove the blocked solvent filter from the bottle-head assembly and place
it in a beaker with concentrated nitric acid (65%) for one hour.
Thoroughly flush the filter with bidistilled water (remove all nitric acid,
some columns can be damaged by nitric acid).
Replace the filter.
CAUTION
Never use the system without solvent filters. This could cause damage to the pump
valves