Agilent Technologies 1100 Series Manual de instrucciones Pagina 167

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 196
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 166
1100 Series Capillary LC System Manual 157
Options 6
Step 3 and 4 have to be done for each of the two flow
cell capillaries, if this connection type is desired.
There is no stop within the union “Top Seal” for plane
capillaries. Therefore one prefixed ferrule is required
at least.
3 Slide the screw, back ferrule and front ferrule onto
the PEEK capillary (see detail for correct direction).
This is for cell side only!
4 Carefully press the capillary into the adjustment
union. Then tight the fitting screw moderately to fix
the ferrule.
5 Appearance of the prefixed ferrule.
NOTE
Note:
The correct torque for prefixing the ferrules in the
“union top-adjust” and for sealing the cell fittings is
0.5 - 0.7 Nm. For the cell screws, the torque adapter
can be used, see Figure 46 on page 155.
back ferrule
screw
front
ferrule
~ 0.3 mm clearance
Vista de pagina 166
1 2 ... 162 163 164 165 166 167 168 169 170 171 172 ... 195 196

Comentarios a estos manuales

Sin comentarios