1100 Series Capillary LC System Manual 173
Performance Specifications 7
For specification on the 500 nl flow cell refer to Table 85 on page 153.
Maximum pressure 50 bar
Control and data
evaluation
Agilent ChemStation for LC
Analog outputs Recorder/integrator: 100 mV or 1 V, output
range 0.001 – 2 AU, two outputs
Communications Controller-area network (CAN), GPIB,
RS-232C, APG Remote: ready, start, stop and
shut-down signals, LAN optional
Safety and maintenance Extensive diagnostics, error detection and display
(through control module and ChemStation), leak
detection, safe leak handling, leak output signal
for shutdown of pumping system. Low voltages in
major maintenance areas.
GLP features Early maintenance feedback (EMF) for continuous
tracking of instrument usage in terms of lamp burn
time with user-settable limits and feedback
messages. Electronic records of maintenance and
errors. Verification of wavelength accuracy with
built-in holmium oxide filter.
Housing All materials recyclable.
* ASTM: “Standard Practice for Variable Wavelength Photometric Detectors Used in Liquid Chroma-
tography”.
Tabl e 92 Performance Specifications Agilent 1100 Series Diode Array Detector
Type Specification Comments