1100 Series Capillary LC System Manual 153
Options 6
500 nl Flow Cell Kit G1315-68714
This section describes the 500 nl flow cell for Agilent 1100 Series diode array
detector and multiple wavelength detector.
Features
small dispersion through:
500 nl, 10 mm pathlength flow cell
novel PEEK jacked quartz capillaries
(inlet 50 µm i.d., outlet 75 µm i.d.)
novel “top sealing” fitting concept
low RI sensitivity for flat baselines at low flow gradients with the use of an
optical reference wavelength
good sensitivity through 10 mm pathlength and acceptable noise level
the cartridge type concept allows customer specific capillary connections
up to the front end of the quartz cell
Performance Specification
Tabl e 85 Performance specification 500 nl flow cell
Type Specification
Pathlength 10 mm
Volume 500 nl
Pressure Operating range 0–5 MPa (0–50 bar, 0–725 psi)
internal diameter of capillaries inlet: 50 µm, outlet: 75 µm
length. of capillaries inlet 400 mm, outlet 700 mm
material of capillaries quartz with PEEK coating
material in contact with solvent quartz, PEEK
Noise specification 2 - 3 of times higher than the 10 mm STD flow cell at
0.05 ml/min