24 1100 Series Capillary LC System Manual
1 Installing your Capillary LC System
Purging the Pump
1 Make sure that the 1/8 inch plastic waste tube is tightly connected to the
barbed waste fitting of the pump EMPV, and routed to an appropriate waste
container.
2 Turn on the LC System. All system parameters should be set to default. The
degasser should also be turned on at this time.
3 Initialize the system. Then, access the pump controls and make sure the
pump mode is set to Normal.
4 Access the pump Purge control. Set up a purge table which will purge all
channels for 5 minutes each, at a flow of 2500 µl/min. Then, start the purge.
When the pump has been turned off for a certain time (for example, overnight), oxygen will re-diffuse into the channels between the degasser and the pump. It is suggested to purge each channel at 2500 ul/min for 1 minute at th e beginning of each day.
NOTE
When the pump has been turned off for a certain time (for example, overnight), oxygen will
re-diffuse into the channels between the degasser and the pump. It is suggested to purge
each channel at 2500 µl/min for 1 minute at the beginning of each day.