Agilent Technologies 1100 Series Manual de servicio Pagina 106

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 318
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 105
106 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
2
Press the release button and open the flow cell door. 3 Disconnect the cell inlet and the cell outlet capillary and
the waste capillary from the capillary holder.
4 Remove the flow cell by pressing the flow cell holder.
Note:
The label attached to the flow cell provides
information on part number, path length and
maximum pressure.
If you want to replace flow cell parts, see “Repairing
the Standard or Semi-Micro Flow Cell” on page 108
or “Repairing the Standard or Semi-Micro Flow
Cell” on page 108.
Vista de pagina 105
1 2 ... 101 102 103 104 105 106 107 108 109 110 111 ... 317 318

Comentarios a estos manuales

Sin comentarios