Agilent Technologies 1100 Series Manual de servicio Pagina 131

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 318
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 130
1100 Series DAD and MWD Reference Manual 131
Repairing the Detector 4
Replacing Leak Handling System Parts
1 Remove the front cover.
2 Pull the leak funnel out of the leak funnel holder.
3 Pull out the leak funnel with the tubing.
4 Insert the leak funnel with the tubing in its position.
5 Insert the leak funnel into the leak funnel holder.
6 Replace the front cover.
When required If the parts are corroded or broken
Tools required None
Parts required Leak funnel 5061-3356
Leak funnel holder 5041-8389
Leak tubing (120 mm) 0890-1711
Figure 32 Replacing Leak Handling System Parts
Leak tubing
Leak funnel
Leak funnel holder