Agilent Technologies 1100 Series Manual de servicio Pagina 169

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 318
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 168
1100 Series DAD and MWD Reference Manual 169
Repairing the Detector 4
16 Replace the leak pan into the cabinet bottom.
Next steps:
Reconnect the leak sensor cable to the processor board
(J21).
Replace the optical unit, see “Installing the Optical Unit
(G1315A/G1365A)” on page 171 or “Installing the
Optical Unit (G1315B/G1365B)” on page 173.
Replace the front cover, top cover and top foam section,
see “Replacing the Foam and the Top Cover” on
page 175.
Replace the detector into the stack.
Reconnect the cables.