Agilent Technologies 1100 Series Manual de servicio Pagina 219

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 318
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 218
1100 Series DAD and MWD Reference Manual 219
Identifying Parts and Materials 5
Agilent 1100 to HP 1090 LC or Signal Distribution Module
Agilent 1100 to General Purpose
Connector
01046-60202
Pin
HP 1090
Pin
Agilent 1100
Signal Name Active
(TTL)
1 1 - White Digital ground
NC 2 - Brown Prepare run Low
4 3 - Gray Start Low
7 4 - Blue Shut down Low
8 5 - Pink Not connected
NC 6 - Yellow Power on High
3 7 - Red Ready High
6 8 - Green Stop Low
NC 9 - Black Start request Low
Connector
01046-60201
Pin
Universal
Pin
Agilent 1100
Signal Name Active
(TTL)
1 - White Digital ground
2 - Brown Prepare run Low
3 - Gray Start Low
4 - Blue Shut down Low
5 - Pink Not connected
6 - Yellow Power on High
7 - Red Ready High
8 - Green Stop Low
9 - Black Start request Low
5 - Key
Vista de pagina 218
1 2 ... 214 215 216 217 218 219 220 221 222 223 224 ... 317 318

Comentarios a estos manuales

Sin comentarios