178 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Assembling the Main Cover
When required If cover is broken
Tools required None
Parts required Plastics kit 5062-8582 (includes base, top, left and right)
NOTE
The plastics kit contains all parts, but it is not assembled.
WARNING
In case you insert the left or right side in the opposite position, you may not be able
to remove the side from the top part.
1 Place the top part on the bench and insert the left and right
side into the top part.
2 Replace the cover.
3 Replace the detector into the stack and reconnect the cables and capillaries.
4 Turn on the detector.