150 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Replacing Fan Parts (G1315B/G1365B)
NOTE
With the introduction of the G1315B/G1365B DAD/MWD a heater and a temperature
sensor was added to the standard fan assembly. Also the air flow direction is reversed
(now: rear to front). Use this procedure for the the G1315B/G1365B only. For
G1315A/G1365A detectors refer to “Exchanging the Fan (G1315A/G1365A)” on page 148.
When required If the fan is defective or noisy, if the temperature sensor or the heater fails
Tools required Screwdriver POZI 1 PT3, a pair of pliers
Parts required Fan assembly 3160-1016,
Temperature sensor assembly G1315-60000,
Heater assembly G1315-60003
Preparations for this procedure:
Turn off the lamp.
Switch off the detector, and disconnect the cables and
capillaries.
Remove detector from stack and place it on the working
bench.
Remove front cover, top cover and top foam section, see
“Removing the Top Cover and Foam” on page 140.
1 Disconnect the connectors from the main board; fan (J16),
temperature sensor (J18) and heater (J17) and pull the fan
assembly out of its location.