Agilent Technologies 1100 Series Manual de servicio Pagina 163

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 318
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 162
1100 Series DAD and MWD Reference Manual 163
Repairing the Detector 4
2 Pull the filter motor off the shaft. Keep the spring in a safe
place.
3 Before installing the filter motor, press the filter lever in
from the other side.
4 While counter-holding the filter lever, press the filter
motor onto the filter shaft until it clicks into the final
position.
Next steps:
Check that the filter lever is moveable.
Reinstall the optical unit, see “Installing the Optical Unit
(G1315A/G1365A)” on page 171 or “Removing the
Optical Unit (G1315B/G1365B)” on page 156.
Reconnect the filter motor cable to connector J18
(G1315/65A) or J19 (G1315/65B) on the detector main
board.
Reinstall the front cover, top cover and top foam section,
see “Replacing the Foam and the Top Cover” on
page 175.
Replace the detector into the stack.
Reconnect the power cable and turn on detector.