1100 Series DAD and MWD Reference Manual 31
How to optmize the Detector 2
Optimization Overview
Table 4 Optimization Overview
Parameter Impact
1 Selection of flow cell
Choose flow cell according to used column, see Figure 7.
peak resolution versus sensitivity
2 Connection of flow cell
For flow rates from 0.5 ml/min connect column using the
zero-dead-volume fittings of the detector.
For small column i.d. (e.g 1 mm) the inlet capillary of the micro flow cell
can be connected directly to the column.
chromatographic resolution
3 Setting the peak width (response time)
Use peak width according Figure 7 as starting point.
Set the peak-width close to the width of a narrow peak of interest in
your chromatogram.
peak resolution versus sensitivity versus
disk space
4 Setting wavelength and bandwidth
Sample wavelength:
Never miss a peak by the use of a browser wavelength like 250 nm
with 100 nm bandwidth.
Select specific wavelength with reduced bandwidth if you need
selectivity, e.g. 250,10 nm and 360,100 nm as reference wavelength.
Set the sample wavelength to a peak or valley in the spectrum to get
best linearity for high concentrations.
Reference wavelength:
Select the reference wavelength with broad bandwidth (30...100 nm)
wavelength range where your analytes have little or no absorbance
(e.g. sample at 254 nm, reference at 320 nm).
sensitivity versus selectivity
sensitivity versus linearity
baseline drift due to RI effects.