140 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Removing the Top Cover and Foam
When required For all repairs inside the detector
Tools required Screwdriver POZI 1 PT3
Screwdriver
Parts required Depends on the work inside and the following procedures
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Disconnect capillaries.
Remove detector from stack and place it on the working
bench.
1 Press the release buttons and remove the front cover.
Unclip the waste funnel assembly.