148 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Exchanging the Fan (G1315A/G1365A)
NOTE
Use this procedure for the the G1315A/G1365A only. For G1315B/G1365B detectors refer
to “Replacing Fan Parts (G1315B/G1365B)” on page 150.
When required If the fan is defective or noisy
Tools required Screwdriver POZI 1 PT3
Parts required Fan assembly 3160-1016
Preparations for this procedure:
Turn off the lamp.
Switch off the detector, and disconnect the cables and
acpillaries.
Remove detector from stack and place it on the working
bench.
Remove front cover, top cover and top foam section, see
“Removing the Top Cover and Foam” on page 140.
1 Disconnect the connectors from the main board (fan J16)
and pull the fan assembly out of its location.