296 1100 Series DAD and MWD Reference Manual
8 Specifications
Performance Specifications
Table 5 0 Performance Specifications Agilent 1100 Series DAD and MWD
Type Specification Comments
Detection type 1024-element photodiode array
Light source Deuterium and tungsten lamps
Wavelength range 190 – 950 nm
Short term noise
(ASTM)
Single and
Multi-Wavelengt
±1×10
-5
AU at 254 and 750 nm See “” on page 297
Drift 2 × 10
-3
AU/hr at 254 nm See “” on page 297
Linear absorbance
range
> 2 AU (upper limit) See “” on page 297
Wavelength accuracy ± 1 nm Self-calibration with deuterium lines,
verification with holmium oxide filter
Wavelength bunching 1 – 400 nm Programmable in steps of 1 nm
Slit width 1, 2, 4 , 8, 16 nm Programmable slit
Diode width < 1 nm
Flow cells Standard: 13 µl volume, 10 mm cell
path length and 120 bar (1760 psi)
pressure maximum
Semi-Micro: 5 µl volume, 6 mm cell
path length and 120 bar (1760 psi)
pressure maximum
High pressure: 1.7 µl volume, 6 mm cell
path length and 400 bar (5880 psi)
pressure maximum
500 nano: 0.5 µl volume, 10 mm cell
path length and 50 bar (725 psi)
pressure maximum
See “Optimization Overview” on
page 31