1100 Series DAD and MWD Reference Manual 113
Repairing the Detector 4
If you want to replace the gasket only, continue with
step 8.
2 Use a tooth pick to remove the quartz window from the
window assembly.
3 If the washers fall out of the window assembly, they must be inserted in the correct order with the Teflon
®
ring to prevent
any leaks from the flow cell window.
Follow the procedure “Repairing the Standard or Semi-Micro Flow Cell” on page 108 for reassembling.
Orientation of Flow Cell Parts
6
1 window screw
contains items 2, 3, 4, 5
and 6
2 seal ring
3 quartz window
4 compression washer
5 spring washers
4 window holder
6 window screw
7 flow cell
8 inlet capillary
9 outlet capillary
5
4
2
3
2
7
1
9
8