Agilent Technologies 1100 Series Manual de servicio Pagina 307

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 318
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 306
1100 Series DAD and MWD Reference Manual 307
Safety Information A
UV-Radiation
Emissions of ultraviolet radiation (200-315 nm) from this product is limited
such that radiant exposure incident upon the unprotected skin or eye of
operator or service personnel is limited to the following TLVs (Threshold
Limit Values) according to the American Conference of Governmental
Industrial Hygienists:
Typically the radiation values are much smaller than these limits:
Table 5 2 UV-Radiation Limits
Exposure/day Effective Irradiance
8 hours 0.1 µW/cm
2
10 minutes 5.0 µW/cm
2
Table 5 3 UV-Radiation Typical Values
Position Effective Irradiance
Lamp installed, 50 cm distance Average 0.016 µW/cm
2
Lamp installed, 50 cm distance Maximum 0.14 µW/cm
2