1100 Series DAD and MWD Reference Manual 81
Troubleshooting and Test Functions 3
Illegal Value From „Air Inlet" Temperature Sensor
This temperature sensor (located on the detector main board) delivered a
value outside the allowed range. The parameter of this event equals the
measured temperature in 1/100 centigrade. As a result the temperature
control is switched off.
Probable Causes
The sensor is defect
Detector is exposed to illegal ambient conditions.
Suggested Actions
Verify that the ambient conditions are within the allowed range.
Exchange detector main board.