1100 Series DAD and MWD Reference Manual 297
Specifications 8
Control and data
evaluation
Agilent ChemStation for LC
Analog outputs Recorder/integrator: 100 mV or 1 V,
output range 0.001 – 2 AU, two
outputs
Communications Controller-area network (CAN), GPIB,
RS-232C, APG Remote: ready, start,
stop and shut-down signals, LAN
optional
Safety and maintenance Extensive diagnostics, error detection
and display (through control module
and ChemStation), leak detection, safe
leak handling, leak output signal for
shutdown of pumping system. Low
voltages in major maintenance areas.
GLP features Early maintenance feedback (EMF) for
continuous tracking of instrument
usage in terms of lamp burn time with
user-settable limits and feedback
messages. Electronic records of
maintenance and errors. Verification of
wavelength accuracy with built-in
holmium oxide filter.
Housing All materials recyclable.
Table 5 0 Performance Specifications Agilent 1100 Series DAD and MWD, continued
Type Specification Comments
NOTE
ASTM: “Standard Practice for Variable Wavelength Photometric Detectors Used in Liquid
Chromatography”.
Reference conditions: cell path length 10 mm, response time 2 s, flow 1 ml/min LC-grade
Methanol, slit width 4 nm.
Linearity measured with caffeine at 265 nm.
For environmental conditions refer to “Environment” on page 15.