16 1100 Series DAD and MWD Reference Manual
1 Installing the Detector
Physical Specifications
Table 1 Physical Specifications
Type Specification Comments
Weight 11.5 kg (26 lbs)
Dimensions
(width × depth × height)
345 × 435 × 140 mm
(13.5 × 17 × 5.5 inches)
Line voltage 100 – 120 or 220 – 240 VAC, ± 10 % Wide-ranging capability
Line frequency 50 or 60 Hz ± 5%
Power consumption (G1315/65B) 300 VA / 125 W / 427 BTU Maximum
Ambient operating temperature 0 – 55 °C (32 – 131 °F) .
Ambient non-operating temperature -40 – 70 °C (-4 – 158 °F)
Humidity < 95%, at 25 – 40 °C (77 – 104 °F) Non-condensing
Operating altitude Up to 2000 m (6500 ft)
Non-operating altitude Up to 4600 m (14950 ft) For storing the detector
Safety standards: IEC, CSA, UL, EN Installation category II, pollution degree 2