Vista de pagina 113
98 Chapter 3
Basic Digital Operation
Using Waveform Markers
13.Press the Enter softkey.
14.Press the
Mode Setup softkey to return to the first-level softkey menu.
Verifying Marker Operation
In this example, you learn how to verify marker operation. If you have not created waveform
segments and applied makers, complete the steps in the previous sections, “Creating
Waveform Segments” on page 77 and “Placing a Marker at the First Point within a
Waveform on page 91.
Once you set a marker on a waveform segment, you can detect the marker pulse at the rear
panel event connector (EVENT 1, 2, 3, or 4) that corresponds to the marker number. (EVENT
1 for this example).
1. Turn on the ARB modulation format or in the dual ARB player, select a waveform and then
turn the dual ARB player on.
Remember a marker is placed on the first point for marker one and two when a waveform
is generated.
2. Connect the oscilloscope input to the EVENT 1 connector, and trigger on the Event 1
signal.
3. Adjust the oscilloscope until the marker pulse is present on the display.