Chapter 3 67
Basic Digital Operation
Arbitrary (ARB) Waveform Header Files
Figure 3-3 ARB Setup Softkey Menu and Marker Utilities
4. Press
ARB Sample Clock > 5 > MHz.
5. Press
More (1 of 2) > Modulator Atten n.nn dB Manual Auto to Manual > 15 > dB.
6. Press
I/Q Mod Filter Manual Auto to Manual > 2.1 MHz.
7. Press
More (2 of 2) > Marker Utilities > Marker Routing > Pulse/RF Blank > Marker 1.
Marker one is now set to blank the RF output at the set marker point(s). See “Using
Waveform Markers” on page 90 for information on setting markers.
8. Press
Return > Marker Polarity > Marker 1 Polarity Neg Pos to Neg.
9. Press
Return > Return > Header Utilities.
Notice that the Current Inst. Settings column now differs from the Saved Header
Dual ARB
Player Softkey
It does not Appear