40 Chapter 2
Basic Operation
Configuring the RF Output
Step sweep provides a linear progression through the start-to-stop frequency and/or
amplitude values. You can toggle the direction of the sweep, up or down. When the
Sweep
Direction Down Up
softkey is set to Up, values are swept from the start frequency/amplitude to
the stop frequency/amplitude. When set to Down, values are swept from the stop
frequency/amplitude to the start frequency/amplitude.
Configuring and Activating a Single Step Sweep
In this procedure, you will create a step sweep with nine, equally spaced points and the
following parameters:
frequency range from 500 MHz to 600 MHz
amplitude from
20 dBm to 0 dBm
dwell time 500 ms at each point
1. Press
Preset.
2. Press
Sweep/List.
This opens a menu of sweep softkeys.
3. Press
Sweep Repeat Single Cont.
This toggles the sweep repeat from continuous to single.
4. Press
Configure Step Sweep.
5. Press Freq Start > 500 > MHz.
This changes the start frequency of the step sweep to 500 MHz.
6. Press
Freq Stop > 600 > MHz.
This changes the stop frequency of the step sweep to 600 MHz.
7. Press
Ampl Start > 20 > dBm.
This changes the amplitude level for the start of the step sweep.
8. Press
Ampl Stop > 0 > dBm.
This changes the amplitude level for the end of the step sweep.
9. Press
# Points > 9 > Enter.
This sets the number of sweep points to nine.
10.Press
Step Dwell > 500 > msec.
This sets the dwell time at each point to 500 milliseconds.