Chapter 2 39
Basic Operation
Configuring the RF Output
4. Press RF On/Off.
The display annunciator has changed from RF OFF to RF ON. The power at the
RF OUTPUT connector is
20 dBm.
5. Press
Incr Set > 10 > dB.
This changes the amplitude increment value to 10 dB.
6. Use the up arrow key to increase the output power by 10 dB.
The AMPLITUDE area displays 10.00 dB, which is the power output by the hardware
(−20 dBm plus 10 dBm) minus the reference power (20 dBm). The power at the
RF OUTPUT connector changes to
10 dBm.
7. Press
Ampl Offset > 10 > dB.
This enters a 10 dB offset. The AMPLITUDE area displays 20.00 dB, which is the power
output by the hardware (
10 dBm) minus the reference power (20 dBm) plus the offset
(10 dB). The OFFS indicator is activated. The power at the RF OUTPUT connector is still
10 dBm.
Configuring a Swept RF Output
The signal generator has two sweep types: step and list.
NOTE List sweep data cannot be saved within an instrument state, but can be saved
to the memory catalog. For instructions on saving list sweep data, see “Storing
Files” on page 55.
During swept RF output, the FREQUENCY and AMPLITUDE areas of the signal
generator’s display are deactivated, depending on what is being swept.
This section provides an explanation of the differences between step sweep and list sweep.
You will learn two ways to configure the signal generator’s RF output to sweep a defined set of
frequency and amplitude points. You will create a step sweep and then you will use these
points as the basis for a new list sweep.
Step Sweep
When a step sweep is activated, the signal generator sweeps the RF output based on the
values entered for RF output start and stop frequencies and amplitudes, a number of equally
spaced points (steps) to dwell upon, and the amount of dwell time at each point. The
frequency, amplitude, or frequency and amplitude of the RF output will sweep from the start
amplitude/frequency to the stop amplitude/frequency, dwelling at equally spaced intervals
defined by the
# Points softkey value.