364 Chapter 15
W-CDMA Uplink Digital Modulation for Receiver Test
Generating a Single PRACH Signal
A default PRACH signal output has a single preamble and a message part with a 20 ms
transmission time interval (TTI). Setting the trigger output to occur on the PRACH pulse lets
the signal analyzer or other measuring equipment that is externally triggered, start its sweep
at the beginning of the PRACH signal.
Generating the Baseband Signal
Press W-CDMA Off On to On.
Setting Up the E4440A PSA for the PRACH Signal
This procedure guides you through setting up the PSA for making a time domain
measurement. Refer to Figure 15-1 on page 348 for connecting the ESG to the PSA. Figure
15-12 on page 365 shows the ESG output signal on the E4440A display.
1. Preset the signal analyzer.
2. Select the W-CDMA mode.
Press
Mode > W-CDMA.
3. Set the center frequency.
Press
Frequency Channel > Center Freq > 1.95 > GHz.
4. Select the time domain measurement mode.
Press
Measure > More 1 of 2 > Waveform (Time Domain).
5. Set a sweep time that shows the components of the PRACH transmission.
Press
Meas Setup > Sweep Time > 50 > ms.
This sets 5 ms as the vertical time divisor (10 divisions x 5 ms = 50 ms)
6. Set the resolution bandwidth for a quick response time that captures the bursting nature
of the PRACH components.
Press
Res BW > 500 > kHz.
7. Trigger the signal analyzer from the rear trigger input using the ESG’s PRACH pulse
trigger signal.
Press
Trig Source > Ext Rear.
8. Zoom in on the RF Envelope display.
Press
Zoom.