Chapter 6 191
BERT Testing
Bit Error Rate Tester–Option UN7
In this example, the triggering sequence is where the trigger delay is active with a cycle
count.
The reference data is generated by stored data bits. If the BER measurement accepts data bits
immediately after receiving a trigger, set the trigger delay to On and the trigger delay count to
a value corresponding to the data format. For PN9 set the delay to 9. If the cycle count is set to
more than 1, it is not necessary to store data bits and no unnecessary delay occurs. Refer to
Figure 6-27 and “Repeat Measurements” on page 193.
Figure 6-27