472 Chapter 16
Troubleshooting
Signal Generator Lock-Up
2. Continue to hold down the Preset key until the following message is displayed:
WARNING You are entering the diagnostics menu which can cause unpredictable
instrument behavior. Are you sure you want to continue?
CAUTION Carefully read the entire message! It may list additional risks with this
procedure.
3. Release the
Preset key.
4. Press
Continue to continue with the sequence (or Abort to abort with no lost files).
At the conclusion of the sequence, follow these steps:
1. Cycle power.
Cycling power restores all previously installed options. You should expect to see several
error messages resulting from calibration files being restored from EEPROM.
2. Perform the DCFM/DC
ΦM calibration.
Refer to the
DCFM/DCΦM Cal softkey description in the Key and Data Field Reference
Volume 1.
3. Agilent Technologies is interested in the circumstances that made it necessary for you to
initiate this procedure. Please contact us at the appropriate telephone number listed in
Table 16-1 on page 474. We would like to help you eliminate any repeat occurrences.