Agilent Technologies ESG Guía de usuario Pagina 194

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 502
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 193
178 Chapter 6
BERT Testing
Measuring RF Loopback BER with Option 300
Figure 6-13
10. Press
Trigger to start the measurement:
After the search is complete, Pass or Fail is displayed on the lower left corner of the
screen, when either of the following occurs:
The result equals or is less than the target percent within the high to low amplitude
range.
The high or low amplitude level is passed at the target% BER/RBER.
Refer to Figure 6-14.
Vista de pagina 193
1 2 ... 189 190 191 192 193 194 195 196 197 198 199 ... 501 502

Comentarios a estos manuales

Sin comentarios