80 Chapter 3
Basic Digital Operation
Using the Dual ARB Player
Playing a Waveform
You can play a waveform sequence or a waveform segment using this procedure. Both
waveform types follow the same process.
In this task, you learn how to play a waveform sequence. If you have not created waveform
segments and used them to build and store a waveform sequence, complete the steps in the
previous procedures, “Creating Waveform Segments” on page 77, Building and Storing a
Waveform Sequence” on page 79.
NOTE Waveform segments must reside in volatile memory to be played individually or
to be played as part of a waveform sequence. Refer to “Loading Waveform
Segments from Non-Volatile Memory” on page 83 for more information.
Selecting a Waveform Sequence
1. Press
Select Waveform.
2. Highlight a waveform (for example, 64CHF+9CHF) from the Sequence column of the
Select Waveform catalog.
3. Press
Select Waveform.
The display shows the currently selected waveform (for example, Selected Waveform:
SEQ:64CHF+9CHF).
Generating the Waveform
Press
ARB Off On until On is highlighted.
This plays the waveform sequence created in the previous sections. During the waveform
sequence generation, the ARB and I/Q annunciators activate. The waveform is now
modulating the RF carrier.
Configuring the RF Output
1. Set the RF carrier frequency.
2. Configure the RF output amplitude.
3. Turn-on the RF output.
The waveform sequence is now available at the signal generator’s RF OUTPUT connector.