Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 173
:MTERegister[:EVENt] (Mask Test Event Event Register)
(see page 798)
Query Syntax
:MTERegister[:EVENt]?
The :MTERegister[:EVENt]? query returns the integer value contained in
the Mask Test Event Event Register and clears the register.
Return Format
<value><NL>
<value> ::= integer in NR1 format.
See Also "Introduction to Root (:) Commands" on page 153
":CHANnel<n>:PROTection" on page 244
":EXTernal:PROTection" on page 273
To MTE bit in
Operation Status
Condition Register
+
OR
:MTEenable
:MTEenable?
Mask Test Event Enable (Mask) Register
1023456798101112131415
:MTERegister[:EVENt]?
Mask Test Event Event Register
Fail
Com-
plete
Auto
Mask
Started
Table 50 Mask Test Event Event Register
Bit Name Description When Set (1 = High = True), Indicates:
15-11 --- --- (Not used.)
10 Auto
Mask
Auto Mask Created Auto mask creation completed.
9 --- --- (Not used.)
8 Started Mask Testing Started Mask testing started.
7-2 --- --- (Not used.)
1 Fail Mask Test Fail The mask test failed.
0Comple
te
Mask Test Complete The mask test is complete.