Vista de pagina 499
500 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
CAN Id specification is set by the :TRIGger:CAN:PATTern:ID
and:TRIGger:CAN:PATTern:ID:MODE commands.
CAN Data specification is set by the :TRIGger:CAN:PATTern:DATA
command.
CAN Data Length Code is set by the :TRIGger:CAN:PATTern:DATA:LENGth
command.
Query Syntax
:TRIGger:CAN:TRIGger?
The :TRIGger:CAN:TRIGger? query returns the current CAN trigger on
condition.
Return Format
<condition><NL>
<condition> ::= {SOF | DATA | ERR | IDD | IDE | IDR | ALL | OVER | ACK}
Errors "- 241, Hardware missing" on page 757
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:MODE" on page 485
":TRIGger:CAN:PATTern:DATA" on page 492
":TRIGger:CAN:PATTern:DATA:LENGth" on page 493
":TRIGger:CAN:PATTern:ID" on page 494
":TRIGger:CAN:PATTern:ID:MODE" on page 495
":TRIGger:CAN:SIGNal:DEFinition" on page 751
":TRIGger:CAN:SOURce" on page 498
NOTE
SOF is the only valid selection for analog oscilloscopes. If the automotive CAN and LIN
serial decode option (Option AMS) has not been licensed, SOF is the only valid selection.