210 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:BUS<n> Commands
Control all oscilloscope functions associated with buses made up of digital
channels. See "Introduction to :BUS<n> Commands" on page 211.
Table 57 :BUS<n> Commands Summary
Command Query Options and Query Returns
:BUS<n>:BIT<m> {{0 |
OFF} | {1 | ON}} (see
page 212)
:BUS<n>:BIT<m>? (see
page 212)
{0 | 1}
<n> ::= 1 or 2; an integer in NR1
format
<m> ::= 0-15; an integer in NR1
format
:BUS<n>:BITS
<channel_list>, {{0 |
OFF} | {1 | ON}} (see
page 213)
:BUS<n>:BITS? (see
page 213)
<channel_list>, {0 | 1}
<channel_list> ::= (@<m>,<m>:<m>
...) where "," is separator and
":" is range
<n> ::= 1 or 2; an integer in NR1
format
<m> ::= 0-15; an integer in NR1
format
:BUS<n>:CLEar (see
page 215)
n/a <n> ::= 1 or 2; an integer in NR1
format
:BUS<n>:DISPlay {{0 |
OFF} | {1 | ON}} (see
page 216)
:BUS<n>:DISPlay? (see
page 216)
{0 | 1}
<n> ::= 1 or 2; an integer in NR1
format
:BUS<n>:LABel
<string> (see
page 217)
:BUS<n>:LABel? (see
page 217)
<string> ::= quoted ASCII string
up to 10 characters
<n> ::= 1 or 2; an integer in NR1
format
:BUS<n>:MASK <mask>
(see page 218)
:BUS<n>:MASK? (see
page 218)
<mask> ::= 32-bit integer in
decimal, <nondecimal>, or
<string>
<nondecimal> ::= #Hnn...n where n
::= {0,..,9 | A,..,F} for
hexadecimal
<nondecimal> ::= #Bnn...n where n
::= {0 | 1} for binary
<string> ::= "0xnn...n" where n
::= {0,..,9 | A,..,F} for
hexadecimal
<n> ::= 1 or 2; an integer in NR1
format