496 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:CAN:SAMPlepoint
(see page 798)
Command Syntax
:TRIGger:CAN:SAMPlepoint <value>
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
The :TRIGger:CAN:SAMPlepoint command sets the point during the bit
time where the bit level is sampled to determine whether the bit is
dominant or recessive. The sample point represents the percentage of time
between the beginning of the bit time to the end of the bit time.
Query Syntax
:TRIGger:CAN:SAMPlepoint?
The :TRIGger:CAN:SAMPlepoint? query returns the current CAN sample
point setting.
Return Format
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:MODE" on page 485
":TRIGger:CAN:TRIGger" on page 499