Agilent Technologies InfiniiVision 6000 Series Manual de servicio Pagina 384

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 934
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 383
384 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:MTESt:DATA
(see page 798)
Command Syntax
:MTESt:DATA <mask>
<mask> ::= binary block data in IEEE 488.2 # format.
The :MTESt:DATA command loads a mask from binary block data.
Query Syntax
:MTESt:DATA?
The :MTESt:DATA? query returns a mask in binary block data format. The
format for the data transmission is the # format defined in the IEEE 488.2
specification.
Return Format
<mask><NL>
<mask> ::= binary block data in IEEE 488.2 # format
See Also ":SAVE:MASK[:STARt]" on page 425
":RECall:MASK[:STARt]" on page 412
Vista de pagina 383
1 2 ... 379 380 381 382 383 384 385 386 387 388 389 ... 933 934

Comentarios a estos manuales

Sin comentarios