Agilent Technologies InfiniiVision 6000 Series Manual de servicio Pagina 283

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 934
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 282
Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 283
:FUNCtion:GOFT:SOURce2
(see page 798)
Command Syntax
:FUNCtion:GOFT:SOURce2 <value>
<value> ::= CHANnel<n>
<n> ::= {{1 | 2} | {3 | 4}} for 4ch models, depending on SOURce1
selection
<n> ::= {1 | 2} for 2ch models
The :FUNCtion:GOFT:SOURce2 command selects the second input channel
for the g(t) source that can be used as the input to the FFT, INTegrate,
DIFFerentiate, or SQRT functions.
If CHANnel1 or CHANnel2 is selected for :FUNCtion:GOFT:SOURce1, the
SOURce2 selection can be CHANnel1 or CHANnel2. Likewise, if CHANnel3
or CHANnel4 is selected for :FUNCtion:GOFT:SOURce1, the SOURce2
selection can be CHANnel3 or CHANnel4.
Query Syntax
:FUNCtion:GOFT:SOURce2?
The :FUNCtion:GOFT:SOURce2? query returns the current selection for the
second input channel for the g(t) source.
Return Format
<value><NL>
<value> ::= CHAN<n>
<n> ::= {{1 | 2} | {3 | 4}} for 4ch models, depending on SOURce1
selection
<n> ::= {1 | 2} for 2ch models
See Also "Introduction to :FUNCtion Commands" on page 278
":FUNCtion:GOFT:SOURce1" on page 282
":FUNCtion:GOFT:OPERation" on page 281
Vista de pagina 282
1 2 ... 278 279 280 281 282 283 284 285 286 287 288 ... 933 934

Comentarios a estos manuales

Sin comentarios