Obsolete and Discontinued Commands 7
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 709
:CHANnel<n>:INPut
(see page 798)
Command Syntax
:CHANnel<n>:INPut <impedance>
<impedance> ::= {ONEMeg | FIFTy}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:INPut command selects the input impedance setting for
the specified channel. The legal values for this command are ONEMeg (1
M
Ω) and FIFTy (50Ω).
Query Syntax
:CHANnel<n>:INPut?
The :CHANnel<n>:INPut? query returns the current input impedance
setting for the specified channel.
Return Format
<impedance value><NL>
<impedance value> ::= {ONEM | FIFT}
NOTE
The :CHANnel<n>:INPut command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :CHANnel<n>:IMPedance command (see page 235)
instead.