Agilent Technologies InfiniiVision 6000 Series Manual de servicio Pagina 531

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 934
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 530
Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 531
:TRIGger:GLITch:GREaterthan
(see page 798)
Command Syntax
:TRIGger:GLITch:GREaterthan <greater_than_time>[<suffix>]
<greater_than_time> ::= floating-point number in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:GLITch:GREaterthan command sets the minimum pulse width
duration for the selected :TRIGger:GLITch:SOURce.
Query Syntax
:TRIGger:GLITch:GREaterthan?
The :TRIGger:GLITch:GREaterthan? query returns the minimum pulse
width duration time for :TRIGger:GLITch:SOURce.
Return Format
<greater_than_time><NL>
<greater_than_time> ::= floating-point number in NR3 format.
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:GLITch:SOURce" on page 537
":TRIGger:GLITch:QUALifier" on page 535
":TRIGger:MODE" on page 485
Vista de pagina 530
1 2 ... 526 527 528 529 530 531 532 533 534 535 536 ... 933 934

Comentarios a estos manuales