Agilent Technologies InfiniiVision 6000 Series Manual de servicio Pagina 175

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 934
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 174
Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 175
:OPEE (Operation Status Enable Register)
(see page 798)
Command Syntax
:OPEE <mask>
<mask> ::= 16-bit integer
The :OPEE command sets a mask in the Operation Status Enable register.
Set any of the following bits to "1" to enable bit 7 in the Status Byte
Register and potentially cause an SRQ (Service Request interrupt to be
generated.
Wa i t
Trig
Run
Arm
Reg
AER?
Run bit set if oscilloscope not stopped
:OPEReration:CONDition?
Operation Status Condition Register
:OPEE
:OPEE?
Operation Status Enable
(Mask) Register
From Overload
Event Registers
To OPER bit in
Status Byte
Register
OVLR
+
OR
1023456798101112131415
Wa i t
Trig
Run
:OPERation[:EVENt]?
Operation Status Event Register
OVLR
HWE
11
5 3
From Hardware
Event Registers
HWE
12
From Mask Test
Event Registers
MTE
MTE
9
Table 51 Operation Status Enable Register (OPEE)
Bit Name Description When Set (1 = High = True), Enables:
15-13 --- --- (Not used.)
12 HWE Hardware Event Event when hardware event occurs.
11 OVLR Overload Event when 50Ω input overload occurs.
10 --- --- (Not used.)
9 MTE Mask Test Event Event when mask test event occurs.
8-6 --- --- (Not used.)
Vista de pagina 174
1 2 ... 170 171 172 173 174 175 176 177 178 179 180 ... 933 934

Comentarios a estos manuales

Sin comentarios