Agilent Technologies InfiniiVision 6000 Series Manual de servicio Pagina 636

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 934
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 635
636 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:WAVeform:BYTeorder
(see page 798)
Command Syntax
:WAVeform:BYTeorder <value>
<value> ::= {LSBFirst | MSBFirst}
The :WAVeform:BYTeorder command sets the output sequence of the WORD
data. The parameter MSBFirst sets the most significant byte to be
transmitted first. The parameter LSBFirst sets the least significant byte to
be transmitted first. This command affects the transmitting sequence only
when :WAVeform:FORMat WORD is selected. The default setting is
LSBFirst.
Query Syntax
:WAVeform:BYTeorder?
The :WAVeform:BYTeorder query returns the current output sequence.
Return Format
<value><NL>
<value> ::= {LSBF | MSBF}
See Also "Introduction to :WAVeform Commands" on page 630
":WAVeform:DATA" on page 638
":WAVeform:FORMat" on page 640
":WAVeform:PREamble" on page 645
Example Code "Example Code" on page 651
"Example Code" on page 646
Vista de pagina 635
1 2 ... 631 632 633 634 635 636 637 638 639 640 641 ... 933 934

Comentarios a estos manuales

Sin comentarios